- page 16

16
PG 8536
Cleaning
Drying
Duration
CW
HW
AD
Energy
Duration
Energy
[mins.]
[l]
[l]
[l]
[kWh]
[mins.]
[kWh]
DES-VAR-TD
54
36.3
24.2
16
3.7
34.3
0.6
DES-VAR-TD AN
57
43
39.5
22
3.7
49.5
0.8
VAR-TD-NR
44
24.3
23.3
16
3.3
34.3
0.5
OPHTHALMOLOGY
48
26.3
36.7
32
2.9
34.3
0.5
ORTHOVARIO
74
34.7
31.3
35
5
34.3
0.4
OXIVARIO
64
38.3
40.7
32
4.4
34.3
0.5
OXIVARIO PLUS
73
27
35
49
4.5
43.3
0.5
SHOE-TD-75/2
27
27.7
35.8
1.5
39.3
0.4
SPECIAL 93/10
48
22
25.5
15.5
3.9
39.3
0.7
CHEM-DESIN
38
26.3
51.7
1.8
40.5
0.4
LAB-STANDARD
33
8.5
38.5
18
2.6
34.3
0.6
LAB-UNIVERSAL
35
8.5
55
21
2.3
34.3
0.6
LAB-INTENSIVE
43
8.5
40
48
2.7
34.3
0.6
LAB-PIPETTES
46
11.5
74.5
44
2.5
34.3
0.3
PLASTICS
38
62.5
20
2.5
44.6
0.4
ORGANICA
41
 1
64
21
2.8
34.3
0.6
ANORGANICA
43
 4
49
48
2.4
34.3
0.6
LAB-OIL
47
 1
80.5
21
2.5
34.3
0.3
Heating: 9 kW (3N AC 400 V, 10.2 kW)
Connection to cold water (15°C), hot water (65°C) and demineralised water (15°C)
Note:
With the exception of OrthoVario, the PG 8535 has the same programmes as the PG 8536.
Cycle times and consumption figures may vary slightly.
Programmes, cycle times, consumption
OPHTHALMOLOGY programme
Miele PG 8535 and PG 8536 washer-
disinfectors now feature a programme
which is specifically designed to cater for
the instruments used in ophthalmological
surgery. By using fully demineralised
water in the final rinse, chemical residue is
reduced to a minimum. This is particularly
important as even the slightest traces of
process chemicals can result in severe
complications, e.g. caustic burns to the
eye. In order to precisely monitor chemical
residue in each stage of the programme,
the use of a PG 8536 with an integrated
conductivity meter is highly recommended
(cf. Page 12 for further details on
conductivity).
Leading instrument manufacturers
recommend Miele reprocessing
systems:
Miele ORTHOVARIO approved for the reprocessing of
current-series Aesculap motor systems.
Recommendation to reprocess ophthalmological instruments
with the Miele system.
Miele OXIVARIO PLUS process approved to prevent the
iatrogenic spread of vCJD
Value-preserving reprocessing of instruments with the VARIO
TD and OXIVARIO processes
Safe reprocessing of flexible endoscopes in models ETD3
and mini ETD2 which are designed and manufactured by
Miele.