- page 9

9
Programmes, cycle times, consumption
VarioTD programme
In the Vario TD programme, pre-cleaning
commences at low temperatures to
prevent the denaturation of blood residues.
After an intensive main wash phase,
thermal disinfection then takes place at
temperatures in excess of 90°C which
are maintained for 5 minutes. In order to
protect surgical instruments, the final rinse
is ideally performed with demineralised
water without any surfactant added.
This programme is suitable for routine
reprocessing according to DIN EN ISO
15883 for all thermally stable instruments.
This approach is particularly gentle on
materials and is recommended for the
reprocessing of transmission instruments.
G7862
Washing / disinfection
Ward Utensils / Surgical Equipment
Duration
CW
WW
Energy
(mins)
(I)
(I)
(kWh)
vario TD AN
47
32.5
40
1.8
vario TD
35
21.5
23.5
1.6
SCAN III
30
21
29
1.4
SCAN II
25
11
29
1.2
SCAN I
20
11
20
1.2
Universal
19
10
29
0.5
free (A)
Heating: 9.0kW (3N AC 400V 9.7kW),
Connection to cold water (15°C) and hot water 65°C)
CW = Cold water, WW = Hot water
Single phase machine - programme duration may vary