70
Washer-disinfectors
PG 8527 (1-door)
PG 8528 (2-door)
Front loader with lift up door
Through feed model with lift up doors
Full glass doors/wash cabinet illumination
For installation singly or in a row
Freshwater system, max. temperature 93°C
Direct connection of mobile units for cleaning/drying hollow instruments
2 circulation pumps [Qmax. l/min]
400/600*
400/600*
Boiler for pre-heating AD water
Controls/Programmes
Profitronic+, 16 Standard programmes
64 programme places
Electrical door lock
Peak load cut-out
Interface for process documentation
Sensor for automatic mobile unit recognition
Spray arm sensor
Conductivity monitor
Water connections
1 x cold water, 2–10 bar pressure (200–1000 kPa) (max. 4°dH)
1 x hot water, 2–10 bar pressure (200–1000 kPa) (max. 4°dH)
1 x AD water, 1.5–10 pressure (200–1000 kPa)
3 inlet hoses
1
/
2
" with
3
/
4
"-threaded union
Drain valve DN 50, odour seal onsite
2 drain pumps DN 22, odour seal onsite
Electrical connection: electrical heating
3 N AC 400 V 50 Hz
Heater wash cabinet [kW]
18
18
Heater boiler [kW]
15
15
Circulation pump [kW]
0.7/1.2*
0.7/1.2*
Total connected load without drying unit [kW]
20
20
Total connected load with electrical drying unit [kW]
20
20
Fuse rating [A]
3 x 32
3 x 32
Electrical connection: steam heating
3 N AC 400 V 50 Hz
Circulation pump [kW]
0.7/1.2*
0.7/1.2*
Total connected load without drying unit [kW]
2
2
Total connected load with steam drying unit [kW]
10
10
Fuse rating [A]
3 x 16
3 x 16
Steam connection G
1
/
2
" (DN 10)
Operating pressure 250–600 kPa (steam drying unit)
Pressurised air connection 600 kPa
Electrical connection: steam/electric convertible heating
3 N AC 400 V 50 Hz
Heater wash cabinet [kW]
18
18
Heater boiler [kW]
15
15
Circulation pump [kW]
0.7/1.2*
0.7/1.2*
Total connected load with electrical drying unit [kW]
20
20
Fuse rating [A]
3 x 32
3 x 32
Steam connection G
1
/
2
" (DN 10)
Operating pressure 250–1000 kPa (electrical drying unit)
Pressurised air connection 600 kPa
Technical data
1...,60,61,62,63,64,65,66,67,68,69 71,72