66
Washer-disinfectors
G 7823 (1-door)
G 7824 (2-door)
Front loader with drop down door
Through feed model with drop down doors
For installation singly or in a row
Freshwater system, max. temperature 93°C
Direct connection of mobile units for cleaning/drying hollow instruments
2 circulation pumps [Qmax. l/min]
300/500*
300/500*
Controls/Programmes
Profitronic, freely programmable
64 programme places
Electrical door lock
Peak load cut-out
Serial interface for process documentation
Sensor for automatic mobile unit recognition
Water connections
1 x cold water, 2–10 bar pressure (200–1000 kPa) 4°dH
1 x hot water, 2–10 bar pressure (200–1000 kPa) 4°dH
1 x AD water, 2–10 pressure (200–1000 kPa)
3 inlet hoses
1
/
2
" with
3
/
4
"-threaded union
Drain valve DN 50, odour seal onsite
2 drain pumps, odour seal onsite, hose DN 22
Electrical connection: electrical heating
3 N AC 400 V 50 Hz
Heater [kW]
9.0
9.0
Circulation pump [kW]
0.4/0.8
0.4/0.8
Total connected load without drying unit [kW]
10.2
10.2
Total connected load with electrical drying unit [kW]
10.2
10.2
Fuse rating [A]
3 x 16
3 x 16
Electrical connection: steam heating
3 N AC 400 V 50 Hz
Circulation pump [kW]
0.4/0.8
0.4/0.8
Total connected load without drying unit [kW]
1.5
1.5
Total connected load with electrical drying unit [kW]
4.5
4.5
Fuse rating [A]
3 x 16
3 x 16
Steam connection G
1
/
2
" (DN 10)
Operating pressure 250–1000 kPa (electrical drying unit)
Pressurised air connection 600 kPa
Electrical connection: steam/electric convertible heating
3 N AC 400 V 50 Hz
Heater [kW]
9.0
9.0
Circulation pump [kW]
0.4/0.8
0.4/0.8
Total connected load with electrical drying unit [kW]
10.2
10.2
Fuse rating [A]
3 x 16
3 x 16
Steam connection G
1
/
2
" (DN 10)
Operating pressure 250–1000 kPa (electrical drying unit)
Pressurised air connection 600 kPa
* 300 l for spray arms / 500 l for direct connection to mobile units
Technical data
1...,56,57,58,59,60,61,62,63,64,65 67,68,69,70,71,72