68
Washer-disinfectors
G 7825 (1-door)
G 7826 (2-door)
Front loader with drop down door
Through feed model with drop down doors
For installation singly or in a row
Freshwater system, max. temperature 93°C
Direct connection of mobile units for cleaning/drying hollow instruments
2 circulation pumps [Qmax. l/min]
300/400*
300/400*
Controls/Programme
Profitronic, freely programmable
64 programme places
Electrical door lock
Peak load cut-out
Serial interface for process documentation
Sensor for automatic mobile unit recognition
Water connections
1 x cold water, 2–10 bar pressure (200–1000 kPa) (max. 4°dH)
1 x hot water, 2–10 bar pressure (200–1000 kPa) (max. 4°dH)
1 x AD water, 2–10 pressure (200–1000 kPa)
3 inlet hoses
1
/
2
" with
3
/
4
"-threaded union
Drain valve DN 50, odour seal onsite
2 drain pumps DN 22, odour seal onsite
Electrical connection: electrical heating
3 N AC 400 V 50 Hz
Heater [kW]
9.0
9.0
Circulation pump [kW]
0.3/0.7*
0.3/0.7*
Total connected load without drying unit [kW]
10.0
10.0
Total connected load with electrical drying unit [kW]
10.0
10.0
Fuse rating [A]
3 x 16
3 x 16
Electrical connection: steam heating
3 N AC 400 V 50 Hz
Circulation pump [kW]
0.3/0.7*
0.3/0.7*
Total connected load without drying unit [kW]
1.65
1.65
Total connected load with electrical drying unit [kW]
9.0
9.0
Fuse rating [A]
3 x 16
3 x 16
Steam connection G
1
/
2
" (DN 15)
Operating pressure 250–1000 kPa (electrical drying unit)
Pressurised air connection 600 kPa
Electrical connection: steam/electric convertible heating
3 N AC 400 V 50 Hz
Heater [kW]
9.0
9.0
Circulation pump [kW]
0.3/0.7
0.3/0.7
Total connected load with electrical drying unit [kW]
10.0
10.0
Fuse rating [A]
3 x 16
3 x 16
Steam connection G
1
/
2
" (DN 15)
Operating pressure 250–1000 kPa (electrical drying unit)
Pressurised air connection 600 kPa
Technical data