46
Technical data
PG 8535 and PG 8536
Washer-disinfectors
PG 8535
PG 8536
Frontloading unit with bottom-hinged door, excl. baskets
Built-under/freestanding unit without lid
Freestanding unit with lid
Freshwater system, max. temperature 93°
Circulation pump [Qmax l/min]
400
600
Controls/programmes
PROFITRONIC+, freely programmable
64 programme slots
Spray arm sensing
Magnetic strip for automatic mobile unit recognition
Conductivity metering
Optional
Network interface for process documentation
Serial printer interface for process documentation
Remote service enabled
Electric door lock
Peak-load negotiation
Water connections
1 x cold water, flow pressure
50 - 1000 kPa
200 - 1000 kPa
1 x cold water for steam condenser, flow pressure
50 - 1000 kPa
200 - 1000 kPa
1 x hot water, flow pressure
50 - 1000 kPa
200 - 1000 kPa
1 x demineralised water, flow pressure
50 - 1000 kPa
50 - 1000 kPa
Feed pump for unpressurised demineralised water
Optional
Optional
4 inlet hoses ½" with ¾" threaded union, (l = approx. 1.7 m)
Drain pump DN 22, head height 100 cm
Steam condenser water drain (DN 22)
WaterProof System (WPS)
Electrical connection
3 N AC 400 V 50 Hz, supply lead approx. 1.7 m, 5 x 2.5 mm²
Heating [kW]
9.0
9.0
Circulation pump [kW]
0.7
1.2
Total rated load [kW]
9.7
10.2
Fuse rating [A]
3 x 16
3 x 16
• = Standard, – = Not available
PG 8535