46
Technical data
PG 8535 and PG 8536
Washer-disinfectors
PG 8535
PG 8536
Frontloading unit with bottom-hinged door, excl. baskets
•
•
Built-under/freestanding unit without lid
•
–
Freestanding unit with lid
–
•
Freshwater system, max. temperature 93°
•
•
Circulation pump [Qmax l/min]
400
600
Controls/programmes
PROFITRONIC+, freely programmable
•
•
64 programme slots
•
•
Spray arm sensing
•
•
Magnetic strip for automatic mobile unit recognition
•
•
Conductivity metering
–
Optional
Network interface for process documentation
•
•
Serial printer interface for process documentation
•
•
Remote service enabled
•
•
Electric door lock
•
•
Peak-load negotiation
•
•
Water connections
1 x cold water, flow pressure
50 - 1000 kPa
200 - 1000 kPa
1 x cold water for steam condenser, flow pressure
50 - 1000 kPa
200 - 1000 kPa
1 x hot water, flow pressure
50 - 1000 kPa
200 - 1000 kPa
1 x demineralised water, flow pressure
50 - 1000 kPa
50 - 1000 kPa
Feed pump for unpressurised demineralised water
Optional
Optional
4 inlet hoses ½" with ¾" threaded union, (l = approx. 1.7 m)
•
•
Drain pump DN 22, head height 100 cm
•
•
Steam condenser water drain (DN 22)
•
•
WaterProof System (WPS)
•
•
Electrical connection
3 N AC 400 V 50 Hz, supply lead approx. 1.7 m, 5 x 2.5 mm²
•
•
Heating [kW]
9.0
9.0
Circulation pump [kW]
0.7
1.2
Total rated load [kW]
9.7
10.2
Fuse rating [A]
3 x 16
3 x 16