44
Technical data
G 7883, G 7893, G 7883 CD
Washer-disinfectors
G 7883
G 7893
G 7883 CD
Frontloading unit with bottom-hinged door, excl. baskets
•
•
•
Freestanding unit with lid, can be built under
•
•
–
Built-under/freestanding unit without lid
–
–
•
Freshwater system, max. temperature 93°
•
•
•
Circulation pump [Qmax l/min]
400
400
400
Controls/programmes
MULTITRONIC Novo Plus/10 programmes
•
•
•
Electric door lock
•
•
•
Buzzer, acoustic signal at end of programme
•
•
•
Programme recontinuation in event of power outage
•
•
•
Serial interface for process documentation,
extendible to USB with PC connection
• (depending on model) •
•
Water connections
1 x cold water, 0.5–10 bar flow pressure (50–1000 kPa)
•
•
•
1 x cold water for steam condenser, 0.5–10 bar flow pressure (50–1000 kPa)
–
•
•
1 x demineralised water, 0.5–10 bar flow pressure (50–1000 kPa)
• (only units without ADP) •
• (only units without ADP)
1 x cold water, 0.5–10 bar flow pressure (50–1000 kPa)
•
•
•
No. of inlet hoses 1/2" with 3/4" threaded union, length approx. 1.7 m
3
4
4
Drain pump DN 22, head height 100 cm
•
•
•
Steam condenser water drain (DN 22)
–
•
•
WaterProof System (WPS)
•
•
•
Electrical connection
, supply lead approx. 1.7 m, 5 x 2.5 mm²
3 N AC 400 V 50 Hz/convertible to 2 N AC 400 V 50 Hz
•/•
•/–
•/–
Heating [kW] (3 N/2 N)
9.0/6.0
9.0/–
9.0/–
Circulation pump [kW] (3 N/2 N)
0.7/0.7
0.7/–
0.7/–
Total rated load [kW] (3 N/2 N)
9.7/6.7
9.7/–
9.7/–
Fuse rating [A] (3 N/2 N)
3 x 16/2 x 16
3 x 16/–
3 x 16/–
Dispenser systems
1 Door dispenser for powder detergent
•
•
•
1 Door dispenser for liquid surfactant
•
•
•
1 DOS 10/30 dispenser pump for liquid acidic agents
•
•
•
1 DOS 60/30 dispenser pump for liquid detergent
–
–
•
Drawer with 2 supply canisters, 5 l each
–
–
•
Connection options
DOS G60, DOS G60/1 or DOS K60, DOS K60/1 for liquid detergent/chemicals
1
1
–
Applications involving oils and greases:
Oil- and grease-resistant gasket
Optional
Optional
Optional
• = Standard, – = Not available
G 7883/G 7893