15
Controls · Programmes · Cycle times
PG 8535, PG 8536
Programmes · Cycle Durations · Energy consumption
PG 8536
Cleaning
Drying*
Cycle Duration CW
HW
AD
Energy
Cycle Duration
Energy
[mins.]
[l]
[l]
[l]
[kWh]
[mins.]
[kWh]
LAB-STANDARD
33
8.5
38.5
18
2.6
34.3
0.6
LAB-UNIVERSAL
35
8.5
55
21
2.3
34.3
0.6
LAB-INTENSIVE
43
8.5
40
48
2.7
34.3
0.6
LAB-PIPETTES
46
11.5
74.5
44
2.5
34.3
0.3
PLASTICS
38
62.5
-
20
2.5
44.6
0.4
ORGANICA
41
1
64
21
2.8
34.3
0.6
ANORGANICA
43
4
49
48
2.4
34.3
0.6
LAB-OIL
47
1
80.5
21
2.5
34.3
0.3
DES-VAR-TD
54
36.3
24.2
16
3.7
34.3
0.6
SPECIAL 93/10
48
22
25.5
15.5
3.97
39.3
0.7
* with 6-minute steam condensation pause (No. 48-50 4.33 mins.)
Heating: 9 kW (3N AC 400 V 10.2 kW)
Connection to cold water (15°), hot water (65°) and demineralised water (15°)
Programmes · Cycle Durations · Energy consumption
PG 8535
Cleaning
Drying
with steam condenser
Cycle Duration CW
HW
AD
Energy
cycle Duration*
Energy
[mins.]
[l]
[l]
[l]
[kWh]
[mins.]
[kWh]
SPECIAL 93°-10'
41
14.7
16.8
11
2.5
39
0.7
DES-VAR-TD
41
23.7**
16.8
11
2.3
34
0.5
SHOE-TD-75/2
24
18.0
23.5
0
1.0
39
0.4
LAB-STANDARD
33
7.3
16.3
22
2.0
34
0.6
ORGANIC
41
2.0
32.0
25
2.1
34
0.6
INORGANIC
39
2.0
32.0
34
1.5
34
0.6
*Programme Duration with steam condensation pause and gentle start
**Cool-down function (CW: +4.5 l/cycle: + 1.75 mins) applies if initial temperature of > 35°C applies
Heating: 9 kW (3N AC 400 V 9.7 kW)
Connection to cold water (15°), hot water (65°) and demineralised water (15°)
Note:
Model PG 8535 features the same programmes as the PG 8536.
Further information on programme durations, etc., on request.
1...,5,6,7,8,9,10,11,12,13,14 16,17,18,19,20,21,22,23,24,25,...48